Loading and Running a Flow Cell on a Oxford Nanopore Technologies MinION
This SOP covers the steps from flow-cell insertion, priming, library loading, starting a sequencing run, and initial checks.
measurementVersion History
Version 1 Current
Effective: 2025-09-17First version. The flow cell FLO-MIN114 (R10.4.1) is referenced. Corresponds to ONT Sequencing / Barcoding Kits V14.
Procedure Steps (Version 1)
Materials and Equipment
Flow cell & device
- MinION device (or compatible device)
- FLO-MIN114 flow cell (R10.4.1)
- Flow cell light shield (for MinION)
Reagents (for loading/priming)
- Flow Cell Flush (FCF)
- Flow Cell Tether (FCT)
- Sequencing Buffer (SB)
- Library Beads (LIB) or Library Solution (LIS)
- Bovine Serum Albumin (BSA) 50 mg/mL (for priming mix) ([Oxford Nanopore Technologies][1])
- Prepared DNA library (from your library prep kit)
- Nuclease-free water
Consumables & equipment
- 1.5 mL Eppendorf DNA LoBind tubes
- Pipettes: P1000, P200, P20, P10 (with filtered tips)
- Microfuge or centrifuge for brief spin-down
- Timer
- Thermal cycler or heat block (if library prep needs incubation)
- Ice bucket
- Optional QC equipment: Qubit fluorometer, Nanodrop, etc.
- Good laboratory practice: gloves, lab coat, sterile tips, aerosol barrier tips
Pre-run Checklist
- Ensure flow cell has been stored and handled according to manufacturer guidelines (e.g. correct temperature).
- Inspect flow cell for shipping damage.
- Perform a Flow Cell Check (via MinKNOW) to verify active pore count. ONT recommends for R10.4.1 flow cells: check before the run. ([Oxford Nanopore Technologies][2])
- Ensure your library prep is completed and you have the correct amount of library (for time-saving you may start priming while doing final library steps).
- Bring flow cell to room temperature for ~20 minutes (per Rapid Barcoding V14 protocol) before loading. ([Oxford Nanopore Technologies][2])
- Have all reagents thawed mixed and on ice as appropriate.
Insert flow cell into device
- Open the lid of the MinION.
- Slide the flow cell under the clip (for the device) and press down firmly on the priming port cover to ensure thermal/electrical contact. ([Oxford Nanopore Technologies][2])
- Ensure the priming port cover is closed until you are ready for priming.
Prepare priming mix
-
Thaw FCF, FCT, SB (and LIB/LIS if required) at room temperature. Then spin down and put on ice. ([Oxford Nanopore Technologies][1])
-
Prepare priming mix with BSA: For FLO-MIN114, combine:
- FCF: 1,170 µL
- BSA 50 mg/mL: 5 µL (final ≈0.2 mg/mL) ([Oxford Nanopore Technologies][1])
- FCT: 30 µL – Mix by pipetting gently. ([Oxford Nanopore Technologies][2])
Remove potential air bubble and prime flow cell
- Open the priming port cover by sliding it clockwise. ([Oxford Nanopore Technologies][2])
- Check for a small air bubble under the priming port cover. Use a P1000 pipette set to ~200 µL, insert tip into priming port, withdraw 20-30 µL (set wheel to ~220-230 µL) until a small volume enters the tip. Ensure continuous buffer covers the pore array. Do not remove more than ~20-30 µL. ([Oxford Nanopore Technologies][1])
- Load 800 µL of the priming mix into the priming port, avoiding air bubbles. Wait 5 minutes. ([Oxford Nanopore Technologies][1])
Prepare library for loading
-
Mix LIB (if using beads) immediately before use (they settle quickly). ([Oxford Nanopore Technologies][1])
-
In a clean 1.5 mL tube, prepare mixture:
- Sequencing Buffer (SB): 37.5 µL ([Oxford Nanopore Technologies][2])
- Library Beads (LIB) or Library Solution (LIS): if using LIB, 25.5 µL; if LIS use 12 µL (per alternative) ([Oxford Nanopore Technologies][2])
- DNA library: 12 µL (or specified amount)
- Total: ~75 µL
-
Mix gently by pipetting up/down just prior to loading.
Final priming and library loading
- Lift the SpotON sample port cover to expose sample port. ([Oxford Nanopore Technologies][1])
- Load 200 µL of the priming mix into the priming port (not the SpotON port) to complete priming. ([Oxford Nanopore Technologies][2])
- Drop-wise load 75 µL of the prepared library into the SpotON sample port. Add each drop, let it flow in before next. ([Oxford Nanopore Technologies][1])
- Replace the SpotON port cover gently, ensure the bung is seated, and close priming port. ([Oxford Nanopore Technologies][2])
- Place the light shield onto the flow cell for best performance. Ensure it sits properly and avoid forcing it under the clip. ([Oxford Nanopore Technologies][1])
- Close the device lid. Flow cell is now ready for sequencing.
Initiate sequencing run in MinKNOW
- Open MinKNOW software and select “Start sequencing”.
- Enter experiment details: sample ID, flow cell model (FLO-MIN114), kit type.
- Select the kit used (e.g., Rapid Sequencing Kit V14 or Rapid Barcoding Kit V14). ([Oxford Nanopore Technologies][2])
- Set output parameters or accept defaults.
- Click “Start”. Monitor pore occupancy, channel activity, yield.
Post-run and Flow Cell Reuse
- If you intend to reuse the flow cell: perform a wash using the Flow Cell Wash Kit (EXP-WSH004) as soon as possible after run completion. Store at +2 °C to +8 °C if not used immediately. ([Oxford Nanopore Technologies][1])
- If returning to vendor: follow manufacturer’s return procedure.
- Record all relevant run metadata: flow cell ID, pore count at start, library details, sample IDs, run duration, yield.
Safety and Handling Notes
- Avoid introducing air bubbles into the pore array—can irreversibly damage pores. ([Oxford Nanopore Technologies][1])
- Always ensure the flow cell matrix is covered by buffer during priming/loading.
- Use filtered tips; avoid rough pipetting which may damage membrane or pores.
- Dispose of used flow cell, tips, reagents according to lab bio-hazard and chemical waste procedures.
- Adhere to manufacturer’s storage and handling instructions for flow cells and reagents.
Troubleshooting – Common Issues
| Issue | Likely Cause | Action |
|---|---|---|
| Low pore occupancy (< 40 %) | Insufficient library loaded, contaminants present | Check library concentration & purity; load recommended amount. (Oxford Nanopore Technologies) |
| Pore occupancy ~0 | No tether added, air bubble introduced | Confirm FCT was added, repeat priming with care. (Oxford Nanopore Technologies) |
| Large number of inactive pores | Air bubble introduced, flow cell mishandled | Review priming steps; avoid rapid loading or pressure on port. (Oxford Nanopore Technologies) |
| “Failed to reach target temperature” in MinKNOW | Device in cold or poorly ventilated location | Move device to room temp, ensure ventilation. (Oxford Nanopore Technologies) |