Fragment Analyzer – HS Large Fragment 50Kb
measurementVersion History
Version 1 Current
Effective: 2025-11-07First version.
Procedure Steps (Version 1)
Method: DNF-464-33 - HS Large Fragment 50Kb.mthds
Instrument: Agilent/Advanced Analytical Fragment Analyzer (33 cm, 10-capillary array)
Analysis Mode: NGS
1. Document Control
- SOP ID: FA-HSLF-50Kb-33cm
- Version: 1.0
- Effective Date: ☐☐☐☐-☐☐-☐☐
- Prepared by: ☐
- Approved by: ☐
- Review cycle: 12 months
2. Purpose
To define the procedure for running high-sensitivity large-fragment DNA analysis on the Fragment Analyzer using the DNF-464-33 method and a 33 cm, 10-capillary array.
3. Scope
Applies to all operators analyzing high molecular weight DNA for NGS library QC or large-fragment assessment using the specified method on a Fragment Analyzer.
4. Responsibilities
- Operator: Execute this SOP, record run metadata, verify QC, and report issues.
- Lab lead: Ensure training, instrument maintenance, and SOP review.
5. Safety
- Wear lab coat, safety glasses, and disposable gloves.
- Handle gels, buffers, and plates per SDS.
- Avoid contact with high voltage areas during operation. Only open instrument when software indicates safe.
- Dispose of consumables as per local regulations.
6. References
- Instrument user manual for the Fragment Analyzer.
- Kit instructions for HS Large Fragment 50Kb chemistry.
- Local laboratory policies for data retention and equipment maintenance.
7. Materials and Equipment
Instrument and configuration
- Fragment Analyzer with 33 cm array, 10 capillaries.
- Array serial and usage count recorded before each run.
Reagents and consumables
- HS Large Fragment 50Kb gel cartridge compatible with
DNF-464-33. - Buffer reservoirs and rinse solution as per kit.
- 96‑well plate(s) and plate seals or caps.
- DNA ladder and markers as required by the kit.
- Nuclease-free water and low-retention tips.
Software
- FA Software v1.2.0.11 or later.
8. Sample Requirements
- DNA in kit-recommended buffer.
- Avoid particulates and salts that interfere with electrokinetic injection.
- Recommended concentration and volume per kit instructions.
9. Pre‑Run Checks
-
Inspect array for leaks, kinks, or salt crystallization. Record array serial and usage count.
-
Confirm instrument details in software:
- Device Serial: record (e.g., 4172)
- FA Version: record (e.g., 1.2.0.11)
-
Verify consumables:
- Correct gel loaded and Gel Selection: Gel 1.
- Fresh buffer and rinse trays in correct positions.
-
Equilibrate reagents to room temperature if required by kit.
-
Prepare plate map with samples, ladder, controls, and blanks.
10. Plate Preparation
- Thaw and mix reagents gently. Spin down briefly.
- Aliquot ladder, controls, and samples into designated wells.
- Seal the plate. Briefly centrifuge to remove bubbles. Keep sealed until loading.
11. Instrument Setup (Software)
Set parameters exactly as follows unless method validation specifies otherwise:
- Method Name:
DNF-464-33 - HS Large Fragment 50Kb.mthds - Analysis Mode:
NGS - Gel Prime:
No - Full Conditioning:
Yes - Gel Prime to Buffer:
No - Gel Selection:
Gel 1 - Perform Prerun:
5.0 kVfor30 s - Rinse: Tray
3, RowA,# Dips: 1 - Marker 1:
No(unless the kit run requires an internal marker; follow kit guidance) - Sample Injection:
5.0 kVfor30 s - Separation:
5.0 kVfor60.0 min - Tray Name:
Tray-1 - # of Capillaries:
10
File management
-
Create or select a run folder. Recommended format:
YYYY MM DD/HH‑MM‑SS/<auto>.raw- Example:
C:\AATI\Data\2022 07 18\14-03-50\2022 07 18 14H 03M.raw
12. Run Procedure
- Load the sealed sample plate into the instrument.
- Confirm tray positions and liquid levels.
- In software, confirm the method parameters match Section 11.
- Start the run. Observe for the first 2–3 minutes for abnormal current, leaks, or errors.
- Allow the 60.0 min separation to complete.
13. Post‑Run Actions
- Save and back up all files generated by the run, including the .raw and analysis reports.
- Export the Run Summary and electropherograms as required by your reporting template.
- If subsequent runs are not planned the same day, follow kit guidance to store gel and perform any required rinses or conditioning before shutdown.
14. Quality Control and Acceptance
- Ladder and control wells exhibit expected profile and sizing (per kit IFU).
- Baseline is stable with distinct peaks where applicable.
- Migration time consistent across capillaries; no persistent current alarms.
- If any criterion fails, document, recondition, and repeat controls before analyzing unknowns.
15. Troubleshooting
- Weak or absent signal: Verify injection settings, sample concentration, plate seal integrity, and capillary tips in wells.
- High noise or drifting baseline: Re‑condition array, check buffer freshness and tray cleanliness.
- Inconsistent sizing: Confirm ladder integrity, plate map accuracy, and temperature stability.
- Frequent current faults: Inspect for bubbles, low liquid levels, or salt contamination. Replace consumables as needed.
16. Maintenance
- Perform Full Conditioning at the start of the day or as required by kit.
- Replace gels and buffers per shelf‑life and use counts.
- Track Array Usage Count at each run and follow vendor guidance for replacement.
- Clean trays weekly or after spills. Inspect electrodes and stage area.
17. Records to Retain
For each run record the following in the run log or ELN:
- Date and time of run.
- Operator name/initials.
- Device Serial and FA Software Version.
- Array Serial # and Usage Count.
- Method Name and all key parameters in Section 11.
- Plate map and sample IDs.
- File path to raw data and reports.
- QC outcomes and any deviations.
18. Run Log Template
| Date | Operator | Device SN | FA Ver | Array SN | Usage Count | Method | Prerun | Injection | Separation | Gel | Rinse | Marker | Capillaries | Data Path | Notes |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| ☐ | ☐ | ☐ | ☐ | ☐ | ☐ | DNF-464-33 | 5.0 kV, 30 s | 5.0 kV, 30 s | 5.0 kV, 60.0 min | Gel 1 | Tray 3 A, 1 dip | No | 10 | C:\…\…\… | ☐ |
B. Abbreviations
- FA: Fragment Analyzer
- HMW: High molecular weight
- IFU: Instructions for Use
- ELN: Electronic Lab Notebook