Nanodrop analysis of nucleic acid concentration and purity
This procedure describes the steps for assessing the concentration and purity of nucleic acids (DNA or RNA) using a Nanodrop spectrophotometer. The analysis measures absorbance at specific wavelengths to determine nucleic acid concentration (ng/µL) and assess purity via the A260/A280 and A260/A230 ratios. Results guide downstream applications by ensuring samples meet quality requirements.
measurementVersion History
Version 1 Current
Effective: 2025-08-08Procedure Steps (Version 1)
Safety
- Wear lab coat, gloves, and eye protection.
- Handle RNA with RNase‑free technique.
- Avoid skin contact with samples and cleaning solutions (e.g., 70% ethanol).
- Dispose of tips/wipes per institutional biosafety guidelines.
Materials and Equipment
- Micro‑volume spectrophotometer (e.g., NanoDrop or equivalent) with nucleic acid method.
- Computer with instrument software and printer (optional).
- Diluent/blank: Nuclease‑free water or low‑EDTA TE (match the sample buffer).
- Lint‑free laboratory wipes; 70% ethanol; DI/NUC‑free water for cleaning.
- Calibrated pipettes and aerosol‑resistant tips.
- Ice rack (optional, to stage samples).
Definitions & Key Formulae
-
Baseline correction (optional): A*λ = Aλ − A320 (or A340, per instrument).
-
Concentration (ng/µL):
- dsDNA: Conc = A*260 × 50 × DF
- RNA: Conc = A*260 × 40 × DF
- ssDNA/oligos: Conc = A*260 × 33 × DF (or use sequence‑specific ε if known) DF = dilution factor used before measurement.
-
Purity ratios (use baseline‑corrected values if applied):
- DNA: A260/A280 ≈ 1.8 (acceptable ~1.7–1.9)
- RNA: A260/A280 ≈ 2.0 (acceptable ~1.9–2.1)
- All: A260/A230 ideally 2.0–2.2 Low A260/A280 suggests protein/phenol; low A260/A230 suggests salts, chaotropes, EDTA, carbohydrates.
Preparation
- Warm‑up & self‑check: Power on per manufacturer guidance; run any required diagnostics if prompted.
- Clean pedestals: Rinse with DI/NUC‑free water → wipe with lint‑free wipe → optional 70% ethanol → water → final wipe.
- Verify blanks match sample buffer: Using water as blank for samples in TE can bias ratios; use the same buffer whenever possible.
- Mix samples gently (pipette up/down or brief vortex) and quick spin to collect.
Procedure — Measurement
-
Launch software and select the nucleic acid method (choose sample type if prompted: dsDNA, RNA, etc.).
-
Baseline/blank:
- Apply 1–2 µL of blank buffer to the lower pedestal.
- Close arm and click Blank.
- Open arm, wipe both pedestals clean.
-
Measure sample:
- Apply 1–2 µL of sample. Ensure it fully bridges the pedestals; avoid bubbles.
- Close arm and click Measure.
- Record displayed A260, A280, A230 (and A320 if shown), concentration, ratios, and DF.
- Open and wipe clean with a fresh lint‑free wipe.
-
Replicates: Measure each sample at least in duplicate (triplicate for critical work). Re‑pipette fresh aliquots for each read.
-
Carryover check: After high‑concentration or problematic samples, run a blank measurement to confirm the reading returns to near zero; re‑clean if not.
-
Dilution if needed: If software flags “over‑range”/non‑linear absorbance or concentration exceeds expected range, dilute the sample with matching buffer, record DF, and re‑measure.
Calculations & Reporting
-
Use software‑reported concentration if baseline correction and sample type are correctly set. Otherwise:
- A*260 = A260 − A320 (if A320 available; else use A260 directly).
- Conc (ng/µL) = A*260 × factor × DF (factor = 50 dsDNA, 40 RNA, 33 ssDNA/approx).
- Ratios: A260/A280 and A260/A230 using corrected absorbances if available.
-
Example (dsDNA): A260=0.325, A280=0.180, A230=0.140, A320=0.005, DF=10 →
- A*260=0.320; A*280=0.175; A*230=0.135 → Conc=0.320×50×10=160 ng/µL;
- A260/A280=1.83; A260/A230=2.37 (acceptable).
Quality Control & Acceptance Criteria
-
Blank check: Post‑cleaning blank absorbance should be ~0.000–0.005 A at 260 nm.
-
Replicate agreement: Concentration replicates within ±5–10%.
-
Purity thresholds (typical, adjust to assay needs):
- DNA acceptable: A260/A280 ≥1.7; A260/A230 ≥1.8.
- RNA acceptable: A260/A280 ≥1.9; A260/A230 ≥1.8.
-
Trend review: Persistently low A260/A230 indicates residual salts/chaotropes—consider re‑purification.
-
Controls (recommended): Measure a known‑concentration reference (DNA/RNA standard) weekly to verify accuracy.
Troubleshooting & Notes
- Low A260/A280 (<1.7 DNA / <1.9 RNA): Protein or phenol carryover. Re‑extract or clean up (e.g., additional wash/ethanol precip/column cleanup).
- Low A260/A230 (<1.8): Salts/guanidine/EDTA/carbohydrates. Additional wash or buffer exchange; ensure blank matches buffer.
- Ratios unusually high (>2.2): Very low ionic strength (e.g., water blank with DNA in water) or baseline drift; measure in TE or enable baseline correction.
- Unstable/erratic readings: Bubbles, insufficient volume, dirty pedestals, or sample not spanning pedestals. Re‑pipette, increase volume slightly, re‑clean.
- Very low concentrations (<~10 ng/µL): Absorbance becomes noise‑limited; confirm with a fluorometric assay (e.g., dye‑based).
- Very high concentrations: Dilute into the same buffer; record DF.
- RNA handling: Use RNase‑free tips/tubes; keep on ice; avoid repeated freeze–thaw.
- pH/ionic strength effects: Ratios shift with pH; TE (pH ~7.5–8.5) gives more stable readings than pure water.
Cleaning & Waste
- After each sample: wipe pedestals with a dry lint‑free wipe.
- Periodically: water → 70% ethanol → water; wipe dry.
- Dispose of tips and wipes in appropriate laboratory waste.
Quick Reference:
- dsDNA: Conc (ng/µL) = A*260 × 50 × DF; purity ~1.8 / >2.0 (A260/A280 / A260/A230).
- RNA: Conc (ng/µL) = A*260 × 40 × DF; purity ~2.0 / >2.0.
- Always blank with the sample’s buffer, enable A320 correction if available, and wipe between every sample.