Qubit™ 1X dsDNA HS Assay (High Sensitivity)
To accurately quantify double‑stranded DNA (dsDNA) in the range of 0.2–100 ng total DNA per tube using the Qubit™ 1X dsDNA HS Assay. Recommended starting sample concentrations are 10 pg/µL to 100 ng/µL. Applicable to research personnel quantifying dsDNA in purified samples, PCR products, and other DNA preparations using Qubit™ 4 or Qubit™ Flex Fluorometers. The assay is selective for dsDNA over RNA and tolerant of common contaminants.
measurementVersion History
Version 1 Current
Effective: 2025-08-25First version.
Procedure Steps (Version 1)
Safety & Handling
- Treat Component A (working solution containing dye) as a potential mutagenic reagent; avoid skin/eye contact and dispose according to local regulations. (p.2)
- Wear appropriate PPE (lab coat, gloves, eye protection).
- Light sensitivity: Protect assay tubes from light; signal in prepared tubes is stable for ≤3 hours at room temperature when protected. (p.1)
Reagents, Consumables & Equipment
Kit (Q33230 – 100 assays; Q33231 – 500 assays):
- Component A: Qubit™ 1X dsDNA HS Working Solution (ready to use) — store 2–8 °C, protect from light.
- Component B: Standard #1 (0 ng/µL in TE) — store 2–8 °C.
- Component C: Standard #2 (10 ng/µL in TE) — store 2–8 °C. When stored as directed, kit stability is ≥6 months from receipt. (p.1, Table 1)
Not provided:
- Qubit™ Assay Tubes, thin‑wall clear 0.5 mL PCR tubes (Cat. Q32856) for Qubit™ 4 or Qubit™ Flex Tube Strips 8 × 200 µL (Cat. Q33252).
- Calibrated pipettes and a P‑2 for 1–2 µL additions; nuclease‑free tips. (pp.2–3)
Instruments:
- Qubit™ 4 or Qubit™ Flex Fluorometer. (Qubit™ 2/3 require the assay program download prior to use.) (p.1)
Critical Parameters & Environmental Conditions
- Assay temperature: Run at room temperature (18–28 °C). Avoid warming tubes by hand or prolonged instrument residence; the fluorometer can heat solutions over minutes.
- Incubation: After mixing standards or samples with working solution, incubate 2 minutes before reading. Signal remains stable ≤3 hours (light‑protected).
- Photostability & re‑reads: Fluorescence drop is <0.3% after 9 reads and <2.5% after 40 reads, but temperature rise inside the instrument can temporarily reduce signal. If re‑reading the same tube, remove it and equilibrate ~30 s at room temperature before the next read.
Preparation (Before You Begin)
- Equilibrate kit components to room temperature.
- Determine whether you will calibrate now or reuse the last calibration. For new users, perform a fresh calibration each run; experienced users may reuse the stored calibration, considering temperature/pipetting conditions. Note the stability limit of prepared standards/samples (≤3 h).
- Plan total working solution required: each tube is 200 µL final volume. Standards require 190 µL working solution + 10 µL standard; samples require (200 µL – sample volume) working solution.
Quick volume guide (per tube): Sample 1 µL → 199 µL working solution; 2 µL → 198 µL; 5 µL → 195 µL; 10 µL → 190 µL; 20 µL → 180 µL. Final volume must be 200 µL.
Tube setup & labeling
- Set out two tubes for standards and one tube per sample (or one strip position per sample for Flex). Use the tube type appropriate for your instrument.
- Label tube lids (not the sides) to avoid interfering with the read. For standards, ensure correct order for calibration.
Load standards and samples
- Add 10 µL of Standard #1 to the “Std1” tube and 10 µL of Standard #2 to the “Std2” tube.
- Add 1–20 µL of each sample to its labeled tube (use a P‑2 for 1–2 µL).
- Add Qubit™ 1X dsDNA HS Working Solution to each tube to reach 200 µL total volume. (Standards: 190 µL; samples: 180–199 µL; consider pre‑aliquoting working solution to avoid cross‑contamination.)
- Vortex 3–5 s to mix. Incubate 2 min at room temperature, protected from light.
Instrument reading
A) Qubit™ Flex Fluorometer 7. Home → 1X dsDNA HS → Read Standards & run samples. 8. Read Standard #1, then Standard #2 (~3 s each). Review the standard graph, then Next. 9. On Insert samples, select active positions; choose units (e.g., ng/µL). Optionally enter kit lot, Tags, Sample IDs. 10. On Sample volume, enter 1–20 µL used per tube (affects assay accuracy range). 11. Insert sample strip, Run samples (~3 s). Results list shows original sample concentration. Add more samples as needed. Optional Molarity/Normalization calculators are available.
B) Qubit™ 4 Fluorometer
7. Home → DNA → 1X dsDNA HS → Read Standards.
8. Read Standard #1, then Standard #2 (~3 s each), review, then Run samples.
9. Select sample volume (1–20 µL) and units on the assay screen.
10. Insert each sample tube and Read tube (~3 s). The top value is original sample concentration; the bottom value is the dilution concentration. Repeat for all samples.
Calibration notes: The instrument fits a curve (modified Hill plot) between the two standards; you may run a new calibration or reuse the previous one (experienced users).
Calculations & Reporting
- The fluorometer automatically converts fluorescence to concentration via the two‑point calibration curve; report the “original sample concentration” produced by the instrument in the selected units.
- Record for each sample: sample ID, input volume, final total volume (200 µL), instrument model/serial, calibration used (new/previous), kit lot number, ambient temperature (if recorded), read time, and result.
Assay Performance, Selectivity & Tolerances
-
Selectivity: Highly selective for dsDNA over RNA
-
Contaminant tolerance: Results typically vary by <10% in the presence of many common contaminants at indicated final assay concentrations—for example:
- NaCl 50 mM (≈1 M in a 10 µL sample), MgCl₂ 1 mM, Sodium acetate 30 mM, Ethanol 1%, Phenol 0.1%, SDS 0.002%, as well as RNA, ssDNA, proteins (BSA/IgG), and dNTPs at levels ~equal to the dsDNA concentration. For best results, add the same contaminant amount to the standards.
-
Dynamic range & confidence: Core detection 0.2–100 ng total DNA per tube with recommended input concentrations 10 pg/µL to 100 ng/µL.
Quality Control
- Standards must read in correct order and generate a plausible calibration curve; if not, repeat calibration with fresh standards.
- Use consistent sample volumes (1–20 µL) across a batch when possible, as this affects the assay accuracy window.
- Avoid temperature drift: do not hold tubes in hand; minimize time in instrument; if re‑reading, equilibrate 30 s at room temperature.
- Document kit lot, instrument, calibration, and incubation time (≥2 min) in the run log.
Limitations
- If sample concentration is outside the HS assay’s accurate range, consider other Qubit™ assays (e.g., dsDNA BR, ssDNA, RNA kits).
- Maximum stability of prepared tubes is 3 h at room temperature when protected from light; read within this window.
Troubleshooting (quick guide)
- Low/variable readings: Verify correct total volume (200 µL), adequate mixing (vortex 3–5 s), and 2‑minute incubation; re‑run calibration; check temperature handling practices.
- Standards fail or curve looks wrong: Replace standards, ensure correct order and volumes; avoid labeling tube sides; recalibrate.
- Signal drift on repeated reads: Remove tube between reads and allow ~30 s equilibration.
Waste Disposal
Dispose of used tips/tubes and any remaining dye‑containing working solution per institutional chemical waste guidelines (treat dye as potentially mutagenic).
References
- **Qubit™ 1X dsDNA HS Assay Kits – User Guide, Pub. No. MAN0017455 Rev. C.0 (08 Dec 2020).